Key concept - Charge envelope. As a single molecule, especially in an ESI source, may pick up multiple charges, the range of mass-to-charge values corresponding to that molecule will vary. These values will often follow a bell-curve distribution over a specific range of values. When the envelope is seen, some simple algebra can be used to determine the range of the charges, and then solve for the mass of the ion. The charge envelope is most often seen in single MS work with larger ions, though it can be applied to a lesser extent to smaller ions as well.
A mass spectra is most commonly displayed with mass-to-charge (m/z, sometimes called "Thompsons") on the x-axis, and relative intensity on the y-axis. The intensity can be though of as being somewhat of a population count for molecules of a given m/z.
Not every mass spectra should be read exacltly the same - different instrument setups will provide differing spectra, as will different molecules. In particular a large molecule (thousands of daltons in mass or more) will give a distinctly different spectra on an ESI-based instrument than a smaller molecule (under 1,000 daltons). Similarly, an ESI instrument will tend to give a different spectra than a MALDI instrument.
For the purpose of this write-up, we will focus primarily on reading the output of ESI instruments, or any ion source that provides multiple charge states.
Reading Single Mass Spectra for Large Molecules
A mass spectra for a large molecule will almost always involve reading a single mass spectrum (as opposed to a tandem or ms^n spectrum). Equally important to consider is the nature of a large molecule, and how it may acquire charge in the ionization process. As discussed earlier, the x-axis of a mass spectra shows the mass-to-charge ratio. This ratio has specific implications for a molecule where the charge is multiple and variable - which occurs frequently with molecules whose masses are several kilodaltons or more. The variable (though integer!) value of charge for members of a single population will result in what is referred to as the charge envelope. The charge envelope for a common protein standard Horse Heart Myoglobin (HHM) is shown below.
You can click on this image to see a larger version of this spectra. The numbers in red on this image are the values of mass-to-charge for ions observed in the instrument. The scale at the bottom of the image shows the m/z range that we were scanning for this molecule - approximately 550 to 2,200. The height of the peaks can be thought of as relative intensity for the population - hence there were more ions observed at 998.37 than at 1060.65.
This image illustrates the concept of a charge envlope, which results from the ionization of a larger molecule in mass spec. These peaks all correspond to the same molecule, just with different charges. As we recall, mass spectrometry observes the mass-to-charge ratio of molecules - hence with charge in the denominator, lower values of m/z correspond to more highly charged molecules. Remember also that charge is quantized as an integer - we'll see the importance of this in a moment.
The charge envelope therefore shows the distribution of charge states for a population of molecules. However, the m/z values themselves - particularly with regards to the value of the charge - also have some additional information. While many large molecules will take up charge such that their charge envelope will be within the range of 0-4,000 m/z, the location within this range is determined somewhat by the accessible surface area of the molecule, and then affected by the ratio of mass to surface area. The HHM molecule whose spectra is shown is largely globular in shape, which means that the total surface area per unit mass is fairly low. Molecules which are more elongated in structure will have more accessible surface area, and hence take on a larger charge for their total mass.
With the spectra obtained for a large molecule, we are still one step away from accurate mass. The next step is to take the charge envlope and solve for the mass of the molecule. This will take advantage of the quantized (integer) nature of charge state, through a process called deconvolution.
Deconvolution can be done manually by taking the numeric value of each peak from the spectra, and entering it into a list. The next step is to then divide the list of peak values by a list of integers, and put the resulting numbers into another list. The integer values need to be iterated by one for each successively smaller value of m/z. For example, if you choose the integer value of "16" for the m/z value of "1060.65", then you would use the integer value of "17" for the m/z value of "998.37". The deconvolution is most accurate when the numbers on the second list are most in agreement.
This image is the result of deconvolution of the earlier spectra of Horse Heart Myoglobin (HHM). You can click on this image to see a larger version of the same image. This deconvolution was done by software, and the tallest peak shows the solved mass from the software that was mathematically "most probable".
The deconvolution graph reads with the units of mass on the x-axis, and intensity (or probability) on the y-axis. When loading a mixture of molecules, it is possible to solve for the components of a mixture if the components are of suitable purity. In the software that was used for this deconvolution, the only input parameters are the range on which to solve the mass, the software will automatically deconvolute to some number of masses to fit the spectra data.
Reading Single Mass Spectra for Small Molecules
Small molecules behave slightly differently in a mass spectrometer than those of larger mass. In particular, a smaller molecule will seldom exhibit the charge envelope that we saw from HHM, at least partially due to the fact that smaller molecules have fewer accessible atoms to acquire additional charge in the mass spectrometer. However, we will see that small molecules still are able to take up multiple charges and give a narrow distribution of m/z values that can be used to determine mass.
The images for this section are based on runs of Glu1-fibrinopeptide B; which will be referred to as "glu-fib" in this write-up. Glu-fib is a readily available peptide that is commonly used as a mass spec standard, we will see in tandem mass spectrometry how we can solve its amino acid sequence.
A generic approach to tandem MS for a compound of good purity is to start with a single MS run, in much the same way that would be done for a larger molecule. The known mass of glu-fib is 1,570.57; the spectra is shown at right (clicking on the spectra will give a larger version of the same image). Much like with the spectra seen previously for HHM, the X axis of this spectra is in units of m/z, and the Y axis is units of intensity. We can see by this spectra that Glu-fib is present in +2 (~785 m/z) and +3 (~531 m/z) and possibly +4 (~324 m/z) species. For tandem MS we will want to choose the most prevalent species with a charge state of 2 or higher (in this case 785, a +2 charge).
A similar deconvolution technique can be applied to this spectra to solve for the mass of the original molecule. However, as the molecule we just saw was a peptide (as opposed to a full-length protein), the mass of the molecule is not as valuable as the information that we will obtain in tandem mass spec on the peptide sequence.
Reading Tandem Mass Spectra for Small Molecules
Now that the single mass spectra has been observed for glu-fib, we can proceed to a tandem mass spec (MS^2) run for the same molecule to observe its component amino acids. On our mass spectrometer we will use argon gas (Ar) to induce the fragmentation of glu-fib to observe its daughter ions.
The hybrid setup of our Q-TOF mass spectrometer allows us to use the quadrupole to both observe and select ions as they pass through the instrument. Having observed the single MS spectra shown earlier for glu-fib, we can now use the quadrupole to select the parent ion species that we want to use for fragmentation. In this case we will select for the m/z equal to 785.
This image is the result of using the quadrupole to filter the mass spectra of glu-fib and select ions that have m/z equal to 785. On this run, no collision gas is used, so the parent ions travel through to TOF without fragmentation. You can click on this image to see a larger version of the same. Notice that none of the other ions from the previous spectra are present in this acquisition; this tells us that the quadrupole is doing its job as a mass filter correctly. If unwanted ions were to pass through the quadrupole and undergo collision in the collision cell, the product tandem spectra would be the product of a mixed population of collision products, which would be much more difficult to interpret.
Once an ion has been selected for collision, we can introduce collision gas to produce daughter ions. The image at right shows the product of selecting for ions at 785 m/z and using argon gas under pressure to fragment the peptide. You can click on this image to see a larger version of the same. Notice that this image does include a peak at 785; this tells us that we likely did not fragment all the ions introduced into the collision cell. In this case a lower flow rate to the ESI source likely could have helped to improve the fragmentation rate.
Notice also that the spectra of the daughter ions shows peaks of both lower and higher m/z values than the selected value of 785. This may seem to be a strange pehonomenon to a novice, as of course these daughter ions should all be of a mass equal to or lower than the parent ion. What could cause the appearance of higher m/z values, then? The answer to this lies in the charge distribution on a molecule. If one were to consider that on this short peptide there are 9 amino acids; each amino acid has roughly equal chance of picking up a proton (and hence a +1 charge) in solution. We are selecting for the +2 species, if each amino acid can pick up only one proton, there are 9^2 = 81 possible charge combinations that could lead to a +2 charge. And remembering that charge is a quantized integer, we know that a daughter ion from a +2 parent can only have a charge of zero, +1, or +2 (we won't be able to detect the zero charge daughter ions). Hence it is possible to have a daughter ion that has most of the mass of the parent, but only half the charge, which will end up with a higher m/z value than the parent. Therefore even though our parent species had a m/z value of 785, we need to scan for ions up to the +1 state ( 785 *2 = 1570) to see all the possible daughter ions.
Chapter 3: Single, Tandem, and ms^n Mass Spectrometry Mass Spectrometry For Dummies Chapter 5: Liquid Chromatography